Abstract
A device has been designed to simultaneously measure the vertical pressure and the anterior-posterior and medial-lateral distributed shearing forces under the plantar surface of the foot. The device uses strain gauge technology and consists of 16 individual transducers (each with a surface area measuring 2.5 x 2.5 cm) arranged in a 4 x 4 array. The sampling frequency is 37 Hz and data may be collected for 2 s. The device was calibrated under both static and dynamic conditions and revealed excellent linearity (±5%), minimal hysteresis (±7.5%), and very good agreement between applied and measured loads (±5%). Vector addition of the distributed loads gave resultant forces that were qualitatively very similar to those obtained from a standard force plate. Data are presented for measurements from the forefoot of 4 diabetic subjects during the initiation of gait, demonstrating that distributed shear and pressure on the sole of the foot can be measured simultaneously. A device has been designed to simultaneously measure the vertical pressure and the anterior-posterior and medial-lateral distributed shearing forces under the plantar surface of the foot. The device uses strain gauge technology and consists of 16 individual transducers (each with a surface area measuring 2.5×2.5 cm) arranged in a 4×4 array. The sampling frequency is 37 Hz and data may be collected for 2 s. The device was calibrated under both static and dynamic conditions and revealed excellent linearity (±5%), minimal hysteresis (±7.5%), and very good agreement between applied and measured loads (±5%). Vector addition of the distributed loads gave resultant forces that were qualitatively very similar to those obtained from a standard force plate. Data are presented for measurements from the forefoot of 4 diabetic subjects during the initiation of gait, demonstrating that distributed shear and pressure on the sole of the foot can be measured simultaneously.
| Original language | English |
|---|---|
| Pages (from-to) | 93-104 |
| Number of pages | 12 |
| Journal | Journal of Applied Biomechanics |
| Volume | 14 |
| Issue number | 1 |
| DOIs | |
| State | Published - Jan 1 1998 |
UN SDGs
This output contributes to the following UN Sustainable Development Goals (SDGs)
-
SDG 3 Good Health and Well-being
Keywords
- Diabetic foot
- Instrumentation
- Plantar pressure
- Shear measurement
Cite this
- APA
- Author
- BIBTEX
- Harvard
- Standard
- RIS
- Vancouver