Eliminating absorbing interference using the H-point standard addition method: Case of Griess assay in the presence of interferent heme enzymes such as NOS

  • Cleveland State University

Research output: Contribution to journalArticlepeer-review

5 Scopus citations

Abstract

Standard calibration methods used to determine trace analytes usually yield significant deviations from the actual analyte value in the presence of interferents in the assay media. These deviations become of particular concern when the concentration of the analyte is low, and when the results are used to draw mechanistic or kinetic conclusions, for instance in enzyme structure-function studies. In these circumstances, the H-point standard addition method (HPSAM) provides superior precision and accuracy. This method is developed here for the case of the spectrophotometric Griess assay used to determine nitrite in various enzymology investigations, such as nitrite determination in studies of nitrite reductases (NiR), or when determining nitrite as a breakdown product of nitric oxide synthesized by NOS enzymes. The results obtained by HPSAM are contrasted with those of the traditional calibration method. © Springer-Verlag 2004.
Original languageEnglish
Pages (from-to)1055-1061
Number of pages7
JournalAnalytical and Bioanalytical Chemistry
Volume379
Issue number7-8
DOIs
StatePublished - Aug 1 2004

Keywords

  • Assay
  • Determination
  • Enzymes
  • Griess
  • Nitrite
  • NOS

Cite this